Valleylab Force FX-C Electrosurgical Unit

  • Fully Refurbished
  • Optional Cart Available
  • Hand & Foot Controls
  • Bipolar & Monopolar

Brochure:

Valleylab Force FX-C Brochure | Valleylab Force FX-C Operators Guide

Description

The Equipped MD Difference.

At Equipped MD we guarantee the best possible pricing on all Medtronic / Valleylab products. We will meet or beat any competitors quote, and offer our “wholesale direct” pricing right out of the gate. This allows us to offer you exceptional value, and custom tailor the pricing to your needs & budget. To learn the details and get in-depth pricing simply hit the request a quote button above or simply give us a call.

REM™ Contact Quality Monitoring System
  • Measurement Frequency: 80 kHz ± 10 kHz
  • Measurement Current: Less than 10 µA
  • Acceptable Resistance Ranges: –REM™ pad – 5-135 ohms –Non-REM™ pad – less than 20 ohms
  • Acceptance range is 5-135 ohms after REM PolyHesive II return electrode is applied.
Adaptive REM™
  • REM™ trip is baseline impedance plus 40%. For example, if the baseline impedance is 30 ohms, the upper level trip approximately 42 ohms. If the pad-patient impedance falls below the baseline impedance, a new baseline is established.
Output Characteristics Bipolar
  • Precise Mode:
    – Maximum P-P Voltage: 450
    – Rated Load (OHMS): 100
    – Maximum Power (Watts): 70
    – Crest Factor* (Typical): 1.5
  • Standard Mode:
    – Maximum P-P Voltage: 320
    – Rated Load (OHMS): 100
    – Maximum Power (Watts): 70
    – Crest Factor* (Typical): 1.5
  • Macro Mode:
    – Maximum P-P Voltage: 750
    – Rated Load (OHMS): 100
    – Maximum Power (Watts): 70
    – Crest Factor* (Typical): 1.5
Monopolar Coag
  • Desiccate Mode:
    – Maximum P-P Voltage: 3500
    – Rated Load (OHMS): 500
    – Maximum Power (Watts): 120
    – Crest Factor* (Typical): 5
  • Fulg. High Crest Factor* Mode:
    – Maximum P-P Voltage: 8500
    – Rated Load (OHMS): 500
    – Maximum Power (Watts): 120
    – Crest Factor* (Typical): 7.0
  • Fulg. Low Crest Factor *Mode:
    – Maximum P-P Voltage: 6900
    – Rated Load (OHMS): 500
    – Maximum Power (Watts): 120
    – Crest Factor* (Typical): 5.5
  • Spray Mode:
    – Maximum P-P Voltage: 9000
    – Rated Load (OHMS): 500
    – Maximum Power (Watts): 120
    – Crest Factor* (Typical): 8
CEM™ Mode
  • Monopolar Cut (Low):
    – Maximum P-P Voltage: 1000
    – Rated Load (OHMS): 300
    – Maximum Power (Watts): 100
    – Crest Factor* (Typical): 1.5
Monopolar Coag (Dessicate):
  • Maximum P-P Voltage: 3500
    – Rated Load (OHMS): 500
    – Maximum Power (Watts): 70
    – Crest Factor* (Typical): 5 * Crest Factor is an indicator of a waveform’s ability to coagulate without cutting
Output Waveforms Bipolar:
  • Precise: 470 kHz sinusoid
  • Standard: 470 kHz sinusoid
  • Macro: 470 kHz sinusoid
Monopolar Cut:
  • Low: 390 kHz sinusoid. Similar to the Pure Cut mode except the maximum voltage is limited to a lower value.
  • Pure: 390 kHz sinusoid
  • Blend: 390 kHz bursts of sinusoid, recurring at 27 kHz intervals. 50% duty cycle envelope.
Monopolar Coag:
  • Desiccate: 240 kHz sinusoid repeated at 39 kHz. 8% duty cycle.
  • Fulgurate: 390 kHz damped sinusoid bursts with a repetition frequency of 30 or 57 kHz into 500 ohms
  • Spray: 390 kHz damped sinusoidal bursts with a randomized repetition centered at 28 kHz. Frequencies include 21 kHz
  • Output power changes by less than 15% or 5 watts, whichever is greater, as the line voltage varies from 104-132 volts and 208- 264 volts (at rated load).
Low Frequency Leakage (50-60 Hz); source current, patient leads, all outputs tied together:
  • Normal polarity, intact chassis ground: <10 µA
  • Normal polarity, ground open: < 50 µA
  • Reverse polarity, ground open: < 50 µA
  • Sink current, 140V applied, all inputs: < 50 µA
PER = 98
  • PER (Power Efficiency Rating) is the measure of an electrosurgical generator’s ability to accurately deliver the selected power into a wide range of tissue types.
Weight and Dimensions
  • Height: 11.1 cm (4-3/8 in.)
  • Width: 35.6 cm (14 in.)
  • Length: 43.9 cm (17 in.)
  • Weight: < 8.1 kg (<18 lbs)
Input Power Requirements
  • Operating range is 85 to 132 AC volts. Maximum current is 7 amperes in Cut and 4 amperes in Coag.
High Frequency Leakage
  • Bipolar: Less than 60 mA (rms)
  • Monopolar: Less than 150 mA (rms)

Additional information

Brand

Medtronic, Covidien, Valleylab

Model

Force FX-C

Title

Go to Top